Preview

Izmeritel`naya Tekhnika

Advanced search
Open Access Open Access  Restricted Access Subscription Access

Noise immunity of pulse wave contour analysis

https://doi.org/10.32446/0368-1025it.2019-3-64-67

Abstract

This article is devoted to analysis of the noise immunity of pulse wave’s contour analysis. Different approaches to the implementation of the pulse wave’s contour analysis were considered. It was demonstrated, that classical approach for the morphological analysis of the pulse wave based on using stiffness index and reflection index do not provide sufficient diagnostic accuracy due to the instability of the maximum detection of the reflected wave. The technique of contour analysis based on spectral transformation of replicated fragments of the pulse wave was proposed. As a diagnostic index the ratio of the amplitudes of the first two harmonics of the spectrum was used. The influence of uncertainty measurement of the pulse wave amplitude on determining the accuracy of the morphological analysis of the pulse wave realized by different methods was researched.

About the Author

A. A. Fedotov
Samara National Research University
Russian Federation


References

1. Expert Consensus Document on arterial stiffness: methodological issues and clinical applications // European Heart Journal. 2006. V. 27 (21). P. 2588-2605.

2. Millasseau S. C., Ritter J. M., Takazawa K., Chowienczyk P. J. Contour analysis of the photoplethysmographic pulse measured at the finger // Hypertension. 2006. V. 8. No. 8. P. 1449-1456.

3. Федотов А. А., Акулов С. А. Измерительные преобразователи биомедицинских сигналов систем клинического мониторинга. М.: Радио и связь, 2013.

4. Weber T., Auer J., O’Rourke M. F., Kvas E., Lassnig E., Berent R., Eber B. Arterial Stiffness, Wave Reflections, and the Risk of Coronary Artery Disease // Circulation. 2004. V. 109. No. 4. P. 184-189.

5. Millasseau S. C., Kelly R. P., Ritter J. M., Chowienczyk P. J. Determination of age-related increases in large artery stiffness by digital pulse contour analysis // Clinical Science. 2002. V. 103. No. 4. P. 371-377.

6. Калакутский Л. И., Федотов А. А. Диагностика дисфункции сосудистого эндотелия методом контурного анализа пульсовой волны // Известия Южного федерального университета. Технические науки. 2009. № 9. С. 93-98.

7. Рангайян Р. М. Анализ биомедицинских сигналов. Практический подход. М.: ФИЗМАТЛИТ, 2007.

8. Федотов А. А. Амплитудно-временно́й метод детектирования характерных точек сигнала пульсовой волны // Медицинская техника. 2012. № 6. С. 22-28.

9. Fedotov A. A., Akulova A. S., Akulov S. A., Reshetnikova M. A. Novel spectral approach for pulse waves contour analysis // IFMBE Proceedings of the 16th Int. Conf. on Biomedical Engineering. 2017. V. 61. P. 16-20.

10. Новицкий П. В., Зограф И. А. Оценка погрешностей результатов измерений. М.: Энергоатомиздат, 1991.


Review

For citations:


Fedotov A.A. Noise immunity of pulse wave contour analysis. Izmeritel`naya Tekhnika. 2019;(3):64-67. (In Russ.) https://doi.org/10.32446/0368-1025it.2019-3-64-67

Views: 113


ISSN 0368-1025 (Print)
ISSN 2949-5237 (Online)