Study of VNIIM quantum voltage standard in the mode of measuring alternating voltage.

https://doi.org/10.32446/0368-1025it.2025-1-113-118

Abstract

In connection with the planned comparisons of AC and DC voltage standards based on the quantum Josephson effect by the International Bureau of Weights and Measures in accordance with the Technical Protocol of Comparisons BIPM.EM-K10, the possibility of a programmable Josephson voltage array from the State Primary Standard of the Unit of Electric Voltage GET 13-2023 for measuring sinusoidal voltages of 0.75 V and 7 V in the frequency range of 62.5–1000.0 Hz was investigated. When measuring voltages in the frequency range of 62.5–200.0 Hz, a digital voltmeter with specially developed software for measuring AC voltage was used together with the quantum voltage standard. The obtained results differ from the measurement results presented in the Technical Protocol of Comparisons. However, the results of the conducted study confirmed the possibility of using a quantum voltage standard based on a programmable Josephson array from GET 13-2023 for comparisons based on the specified Technical Protocol. The interaction of the quantum voltage standard and the AC voltage source was analyzed, and a conclusion was made on the need to control the reproduced AC voltage with a digital voltmeter with the software recommended in the Technical Protocol for comparisons.

About the Authors

A. S. Katkov
D. I. Mendeleyev Institute for Metrology
Russian Federation

Alexander S. Katkov



G. B. Gubler
D. I. Mendeleyev Institute for Metrology
Russian Federation

Gleb B. Gubler



V. I. Shevtsov
D. I. Mendeleyev Institute for Metrology
Russian Federation

Vladimir I. Shevtsov



I. A. Sladovskiy
D. I. Mendeleyev Institute for Metrology
Russian Federation

Ivan A. Sladovskiy



References

1. Josephson B. D. Possible new effects in superconductive tunnelling. Physics Letters, 1(7), 251–253 (1962). https://doi.org/10.1016/0031-9163(62)91369-0

2. Krzymovsky V. I., Katkov A. S. Increasing the accuracy of maintaining the unit of EMF. All-Union Conference “Quantum Metrology and Fundamental Physical Constants”. Abstracts of the report. VNIIM, Leningrad, pp. 173–174 (1982). (In Russ.)

3. Koltik E. D., Galakhova O. P., Krzymovsky V. I. et al. State primary standard of the unit of EMF and the transition to a new size of the volt. Izmeritel’naya Tekhnika, (10), 6–7 (1991). (In Russ.)

4. Katkov A. S., Koltik E. D., Krzhimovskii V. I. et al. Common quantum voltage standard for direct and alternating currents. Development concept. Measurement Techniques, 40(12), 1191–1196 (1997). https://doi.org/10.1007/BF02504165

5. Krutikov V. N., Katkov A. S. Standard of the Volt of Russia. Control and Measuring Instruments and Systems, (3), 25 (2000). (In Russ.)

6. Aleksandrov V. S., Katkov A. S., Telitchenko G. P. New State primary standard and state test scheme for instruments for measuring DC electrical voltage and electromotive force. Measurement Techniques, 45(3), 228–232 (2002). https://doi.org/10.1023/A:1015954501462

7. Katkov A. S., Shevtsov V. I., Gromova J. A. State primary voltage standard of the fourth generation. Measurement Techniques, 66(7), 447–453 (2023). https://doi.org/10.1007/s11018-023-02246-1

8. Solve S., Kim M. -S., Palafox L. et al. Update on the future BIPM on-site comparison program for Josephson ac voltage standards. Conference on Precision Electromagnetic Measurements (CPEM 2020), Denver, USA, pp. 1–2 (2020). https://doi.org/10.1109/CPEM49742.2020.9191727

9. Lee J., Behr R., Katkov A., Palafox L. Error contributions in stepwise synthesized Josephson sine waves. Conference Digest (CPEM 2008). Broomfield, pp. 78–79 (2008).

10. Lee Jinni, Behr Ralf, Palafox Luis et al. An ac quantum voltmeter based on a 10 V programmable Josephson array. Metrologia, 50(6), 612–622 (2013). https://doi.org/10.1088/0026-1394/50/6/612

11. Katkov A. S., Lovtsyus V. E., Bykov A. I. Shevtsov V. I., Novoderezhkin G. V. Reproduction of the Volt Based on SIS and SNS-type Josephson Junctions. Measurement Techniques, 60(6), 589–593 (2017). https://doi.org/10.1007/s11018-017-1240-1

12. Katkov A., Shevtsov V., Gubler G. VNIIM 10 V Programmable Josephson Voltage Standard. Conference on Precision Electromagnetic Measurements (CPEM 2018), Paris, France, pp. 1–2 (2018). https://doi.org/10.1109/CPEM.2018.8500892

13. Gubler G. B., Shapiro E. Z. Implementation of sampling measurement system for new VNIIM power standard. Conference on Precision electromagnetic Measurements (CPEM 2012), Washington, USA, pp. 294–295 (2012). https://doi.org/10.1109/CPEM.2012.6250918

14. Telitchenko G. P., Shevtsov V. I. The new state special primary standard for the unit of electrical potential (the volt) in the frequency range 10–3•107 Hz. Measurement Techniques, 52(7), 687–690 (2009). https://doi.org/10.1007/s11018-009-9334-z


Review

For citations:


Katkov A.S., Gubler G.B., Shevtsov V.I., Sladovskiy I.A. Study of VNIIM quantum voltage standard in the mode of measuring alternating voltage. Izmeritel`naya Tekhnika. 2025;74(1):113-118. (In Russ.) https://doi.org/10.32446/0368-1025it.2025-1-113-118