Preview

Izmeritel`naya Tekhnika

Advanced search
Open Access Open Access  Restricted Access Subscription Access

Measurement of spectral irradiance of deuterium lamps in the wavelength range from 200 nm to 400 nm traceable to high-temperature blackbody

https://doi.org/10.32446/0368-1025it.2018-11-46-51

Abstract

The paper presents the VNIIOFI procedure of measurement of spectral irradiance of deuterium lamps in the wavelength range from 200 nm to 400 nm. Expanded uncertainty of the measurements varies from 1,6 to 6,9 %. The results of VNIIOFI measurements were confirmed in the comparison with PTB. The deviation between two laboratories was not higher than 2 %.

About the Authors

B. B. Khlevnoy
The All-Russian Research Institute for Optical and Physical Measurements Federal State Unitary Enterprise
Russian Federation


M. V. Solodilov
The All-Russian Research Institute for Optical and Physical Measurements Federal State Unitary Enterprise
Russian Federation


References

1. Sperfeld P. Final report on the CIPM key comparison CCPR-K1.b: Spectral irradiance 200 nm to 350 nmю // Metrologia. 2008. V. 45. Technical Supplement. 02002. DOI: 10.1088/0026-1394/45/1A/02002.

2. Гаврилов В. Р., Григорьева И. А., Ивашин Е. А., Отряскин Д. А., Солодилов М. В., Солодилов Е. В., Хлевной Б.Б. Государственный первичный эталон единиц радиометрических и спектрорадиометрических величин в диапазоне длин волн от 0,2 до 25,0 мкм ГЭТ 86-2017 // Измерительная техника. 2018. № хх. C. хх-хх.

3. Sapritsky V. I., Khlevnoy B. B., Khromchenko V. B., Lisiansky B. E., Mekhontsev S. N., Melenevsky U. A., Morozova S. P., Prokhorov A. V., Samoilov L. N., Shapoval V. I., Sudarev K. A., Zelener M. F. Precision blackbody sources for radiometric standards // Appl. Opt. 1997. V. 36. No. 22. P. 5403-5408.

4. Хлевной Б. Б. Метрологические исследования эталонных пирографитовых высокотемпeратурных чёрных тел // Измерительная техника. 2001. № 12. C. 26-30.

5. Prokhorov A. Effective emissivities of isothermal blackbody cavities calculated by the Monte Carlo method using the three-component BRDF Model // Appl. Opt. 2012. V. 51. No. 13. P. 2322-2332.

6. Samoylov M. L., Ogarev S. A., Khlevnoy B. B., Khromchenko V. B., Mekhontsev S. N., Sapritsky V. I. High Accuracy Radiation TSP-type Thermometers for Radiometric Scale Realization in the Temperature Range from 600 to 3200 °C // Temperature: It´s Measurement and Control in Science and Industry: Proc. 8th Symposium on Temperature. 2003. V. 7. P. 583-588.

7. Woolliams E. R., Anhalt K., Ballico M., Bloembergen P., Bourson F., Briaudeau S., Campos J., Cox M. G., Campo D., Dong W., Dury M. R., Gavrilov V., Grigoryeva I., Hernanz M. L., Jahan F., Khlevnoy B., Khromchenko V., Lowe D. H., Lu X., Machin G., Mantilla J. M., Martin M. J., McEvoy H. C., Rougié B., Sadli M., Salim S. G. R., Sasajima N., Taubert D. R., Todd A. D. W., Van den Bossche R., Van der Ham E., Wang T., Whittam A., Wilthan B., Woods D. J., Woodward J. T., Yamada Y., Yamaguchi Y., Yoon H. W. and Yuan Z. Thermodynamic temperature assignment to the point of inflection of the melting curve of high-temperature fixed points // Philos. Trans. Royal Soc. A. 2016. V. 374. No. 2064. P. 20150044.

8. Khlevnoy B. B., Grigoryeva I. A. Long-Term Stability of WC-C Peritectic Fixed Point // Int. J. Thermophys. 2015. No.36. P. 367-373.

9. Birch K. P., Downs M. J. Correction to the updated Edlén equation for the refractive index of air // Metrologia. 1994. V. 31. No. 4. P. 315-316.

10. Sperfeld P., Pape S., Khlevnoy B., Burdakin A. Performance limitations of carbon-cavity blackbodies due to absorption bands at the highest temperatures // Metrologia. 2009. No. 46. S170-173.


Review

For citations:


Khlevnoy B.B., Solodilov M.V. Measurement of spectral irradiance of deuterium lamps in the wavelength range from 200 nm to 400 nm traceable to high-temperature blackbody. Izmeritel`naya Tekhnika. 2018;(11):46-51. (In Russ.) https://doi.org/10.32446/0368-1025it.2018-11-46-51

Views: 108


ISSN 0368-1025 (Print)
ISSN 2949-5237 (Online)